SVAN 948
4 Channel Sound & Vibration Data Logger
|
|
 |
OVERVIEW
The SVAN 948 is all digital, 4
channels, Type 1 vibration and sound level meter and analyser. It is an
ideal choice for the “Human Vibration” and noise measurements in the
occupational health and safety monitoring tasks.
All required weighting filters,
as well as transducer (including triaxial seat accelerometer) are available
with this unit.
Four vibration and/or sound
channels can work in parallel with independently set-up transducers, filters
and RMS detectors time constant (e.g. simultaneous measurement of the Whole
Body and noise dose). Large 32 MB internal memory of the SVAN 948 can
store time history for all channels over the whole working day.
Measurement results are easily
downloaded to any PC using the USB 1.1 interface and SvanPC software.
The SVAN 948, using computational power of its built-in digital signal
processor, can optionally perform 4 channels 1/1 or 1/3 octave and real time
FFT analysis.
The SVAN 948 can be used in hard
environmental conditions, also in the field of the machinery condition
monitoring. Robust, lightweight construction fulfils the exceptional
features of this instrument.
Top
FEATURES
- Four channels simultaneous vibration
and/or sound measurement
- Human Vibration measurements Type 1
accuracy (ISO 8041) including VDV and MTVV
- Noise measurements Type 1 IEC 61672 (IEC
651 and IEC 804)
- Acoustic dose meter function (optional)
- 32 MB internal buffer for logging more
then two weeks of 1 sec results
- 1/1 and 1/3 octave real time analysis
(optional)
- FFT analysis (optional)
- Easy operation
- USB 1.1 interface
- Battery operational time > 8 h
- Handheld, robust case
- Light weight (ca 500 grams only)
Top
For a PDF version of this document please click
here.
TECHNICAL
SPECIFICATIONS
| Sound level meter |
Type 1: IEC 651, IEC 804 and IEC
61672-1 |
| |
Spl, Leq,
SEL, Lden, Ltm3, Ltm5, statistics: Ln (L1-L99), Lmax, Lmin , LPeak, Time
History |
|
Vibration meter |
Type 1: ISO 8041,
RMS, VDV, Peak, Peak-Peak, Max (MTVV), Time History |
| Sound
& Vibration Analyser |
Real
time 1/1 & 1/3 octave or FFT analysis (optional) |
| |
1920-line FFT spectra
calculated in the real time for 20 kHz and sub-bands (optional) |
| Input |
4
channels (channels 1 - 3 – Lemo 4 pin connector and channel 4 – TNC
connector) with IEPE power supply for the accelerometers or microphone
preamplifiers |
|
Accelerometers |
SV 39A/L
seat accelerometer for Whole Body measurements
SV 3023M2 triaxial accelerometer for Hand-Arm measurements
Other IEPE accelerometers optional |
|
Microphones |
SV 22
prepolarised 1/2” condenser microphone (50 mV/Pa sensitivity) with SV 12
preamplifier, or
SV MI17 (Type 2) 1/4” electret microphone |
| Measurement range |
0.003ms-2
- 1000 ms-2- RMS (with 10 mV/ms-2- accelerometer)
or
24dBA RMS - 140 dBA Peak (with 50 mV/Pa microphone) |
|
Internal noise level |
< 17 dBA |
| Dynamic range |
100 dB |
| |
A/D conversion:
4 x 20 bits |
|
Frequency range |
0.5 Hz -
20 kHz (input transducer dependent) |
| |
Sampling
rate: 51.2 kHz |
|
Weighting filters |
W-Bxy,
W-Bz, W-Bc, H-A (ISO 8041),
Wk, Wc, Wd, Wj, Wh (ISO 2631-1 and ISO 5349),
A, C and Lin (Z) (Type 1: IEC 651, IEC 804 and IEC 61672-1) |
| RMS & RMQ detector |
Digital True RMS & RMQ with Peak detection, resolution 0.1 dB,
Integration time programmable up to 24 h, |
| Time constants |
100 ms
to 10 s in the Vibration Meter mode
Slow, Fast, Impulse in SLM mode |
| Display |
LCD 97x32 pixels plus icons
with backlight |
| Memory |
32 MB, non-volatile
(Flash type) |
| Interfaces |
USB 1.1 |
| Power supply |
4 x AA batteries operating time > 8 h |
| |
External power supply 6-24 VDC
/ 250 mA |
| |
USB
interface: 5V@300mA |
| Environmental conditions |
Temperature from -10
˚C to 50
˚C |
| |
Humidity up to 90% RH, non condensed |
| Dimensions |
135 x 82 x 42 mm |
| Weight |
approx. 0.5 kg with batteries |